Welch Allyn AED 10

The Welch Allyn AED 10 is a safe, easy-to-use defibrillation device. The unit is lightweight and mobile and can be used in situations where there could be several minutes before the arrival of advanced life support (ALS) personnel. The AED 10 recognizes ventricular fibrillation and other ventricular tachycardia and guides operators through the defibrillation process. The Welch Allyn AED 10 delivers the defibrillation shock through two self-adhesive, pre-gelled, low-impedance electrode defibrillator pads.

Worldwide Sales & Service

  • Upto 50% Below OEM pricing
  • Reliable Equipment’s for your Facility

Welch Allyn AED 10 Features

  • Two-button operation
  • Extensive voice and visual prompts for the operator
  • Continuous event recording for reporting each use to a printer or computer
  • Weekly self-test to ensure readiness
  • Biphasic energy output
  • Lock-out protection to prevent inadvertent defibrillation
  • Continuous surveillance of battery level

Welch Allyn AED 10 Specifications

Dimensions

  • Height: 2.75” (7 cm)
  • Width: 6.88” (17.5 cm)
  • Depth: 8.25” (21 cm)
  • Weight: 3.5 lbs (1.6 kg)

Data Management

  • Event Documentation: Internal
  • Internal Memory Capacity: 1MB: 250 4-sec ECG samples or 3000 time-stamped events
  • Quick Report: Treatment Summary, Event Log, Test Log
  • Communication: RS-232 via Infrared (IR) port to PC or printer

Defibrillator

  • Output: MRL Orbital Biphasic Truncated exponential
  • Energy Sequence (user configurable): Shock 1: 150 J, 200 J; Shock 2: 150 J, 200 J, 300 J; Shock 3: 150 J, 200 J, 300 J, 360 J
  • Charge Time from “Shock Advised”: 4 sec. to 150 J; 8 sec. to 200 J; 15 sec. to 360 J
  • Analysis Time: 4-16 sec.
  • Combined Analysis and Charge Time From start of analysis to shock ready:
    • from power up: Less than 20 seconds to 200 J
    • after 6 shocks: Less than 30 seconds to 360 J
  • Audible Prompts: 18 audible prompts
  • Visual Prompts: 20 text screen prompts
  • Controls: Four buttons – On/Off, Shock, and 2 software-configurable buttons

Display

  • Type: Backlit Monochrome LCD
  • Size: 2.1” x 1.42” (54 x 36 mm)
  • Resolution: 160 x 100
  • Freq Response: 1 to 40 Hz
  • Low Battery Indicator: Battery Icon gauge on display with 10 capacity levels
  • Backlight: LED

Battery

  • Type: Non-rechargeable, LiMnO2 12V, 2.6 Ah
  • Capacity: 90 discharges at 360 Joules or; 125 discharges at 200 Joules or; 150 discharges at 150 Joules or; 5 hours ECG monitoring
  • Shelf Life (25°C ± 15°C): 10 years (5 years storage + 5 years standby); 5 years standby (after installation).

Additional information

Manufacturer

Welch Allyn