Physio Control Lifepak 1000 Features
- Rugged Construction – Rigorously drop-tested device and protective case and bumpers
- Vehicle Friendly – Designed to ride along in any vehicle without damage from continuous vibrations and other movement
- 360 Joules – Can escalate defibrillation power to an industry leading 360J
- cprMAX Technology – Minimizes CPR interruptions by allowing compressions to continue during AED charging
- ECG Capability – 3-lead ECG function is available when needed
- Shock Counter – Digitally records and displays delivered shocks for added insight
Physio Control Lifepak 1000 Specifications
Display
- Size (Active viewing area)
- 120 mm (4.7 in.) x 89 mm (3.5 in)
- Display Type 320 dot x 240 dot LCD with backlight
- Frequency Response 0.55 Hz to 21 Hz (-3 dB), nominal
- Waveform Sweep Speed :25 mm/sec for ECG, nominal
- Waveform viewing time: Minimum 4 seconds
- Waveform Amplitude 1 cm/mV, nominal
- Display Range Differential: ±1.4 mV full scale, nominal
- Heart Rate 20 to 300 BPM digital display
- Display “—” if heart rate is less than 20 BPM
- Heart symbol flashes for each QRS detection
- Displayed ECG ECG information is received from therapy pads in anterior-lateral or anterior-posterior positions, or from the 3-wire ECG cable in Lead II
ECG Mode
- ECG Display Provides nondiagnostic ECG display of the patient’s heart rhythm.
Manual Mode
- Energy Sequence Delivers energy at levels selected in Setup mode
- Charge Time Charge time:
- 200 joules in less than 7 sec (typical)
- 360 joules in less than 12 sec (typical)
AED Mode
- Shock Advisory System (SAS): ECG Analysis system that advises whether a shock is appropriate, meets rhythm recognition criteria specified in DF80 and IEC 60601-2-4. In AED mode, the device allows a shock only if SAS advises defibrillation. Shock Ready Time Time to first shock (electrodes connected to a patient at power on and initial rhythm finding is Shock Advised):
- Less than 25 sec to 200 joules
- Less than 30 sec to 360 joules
- Energy Sequence: Multiple levels, configurable from 150 to 360 joules
- Shock-to-Shock cycle time (200J to 300J):
- Less than 25 seconds
- Time for a 3-shock sequence (200J/300J/360J):
- Less than 70 seconds
With Infant/Child Pads
- The following specifications apply from 25 to 175 ohms
- Energy Accuracy (into 50 ohms):
- Selected energy ÷ 4 +/- 15%; 86 joules +/- 15% maximum
With Adult Pads
- Patient Impedance Range: 10 – 300 ohms
- The following specifications apply from 25 to 175 ohms.
Energy Accuracy
- 10% of the energy setting into 50 ohms
- 15% of the rated energy output into 25 – 175 ohms
Waveform
- Biphasic Truncated Exponential with voltage and duration compensation for patient impedance.