Philips Heartstart FR3 Features
Built to endure and designed for flexibility
The FR3 is built for harsh environments. It is tested to stringent military standards, meets IP55 rating for protection against dust and jetting water and passes a 1,100 pound crush test. It’s rechargeable battery option provides up to 100 shocks and 3 hours of monitoring time. Plus, FR3 is a highly flexible platform so you can configure it for a wide range of current and evolving protocols.
Rich clinical information to drive patient care
The FR3 delivers patient-specific guidance for the most appropiate initial therapy. For non-cardiac arrest patients, using the 3-Lead ECG cable helps you assess their condition. You get the information you need for confident decision-making. Plus, Philips add-on data management solutions help support a culture of continuous improvement and excellence.
- Automatically powers on by opening the Philips HeartStart FR3 carry case so you can focus on pad placement from the start
- Infant/Child Key automatically decrease the defibrillation therapy and implement the configured infant/child CPR protocols
- Smallest and lightest professional-grade AED among leading global manufacturers
- Bright, high-resolution color LCD makes the FR3 easier to use in a noisy environment
- Pre-connected peel and place SMART Pads III with no foil pouch to reduce deployment time
- 5-Year Factory Warranty from Philips Healthcare.
Philips HeartStart FR3 Specifications
Size with external paddles
- Height: 13.4″ (34 cm)
- Width: 12.4″ (31.3 cm)
- Depth: 7.7″ (19.5 cm)
Size without external paddles
- Height: 11.7″ (29.5 cm)
- Width: 12.4″ (31.3 cm)
- Depth: 7.7″ (19.5 cm)
Weight
- 13.2 lbs. (6 kg): base unit with 1 battery/pads and pads cable.
- 13 lbs. (5.9 kg) with optional 75 mm strip chart printer.
- Paddle tray and external standard paddles add less than 2.5 lbs. (1.1 kg).
- Carrying case adds 4.1 lbs. (1.86 kg). lbs/kg
Defibrillation
- Waveform: Truncated Exponential Biphasic.Waveform parameters adjusted as a function of patient impedance.
- Output Energy: Manual (selected): 1-10/15/20/30/50/70/100/120/150/170/200 Joules into a 50 Ohm load
- AED Mode (single energy output): 150 Joules into a 50 ohm load. Joules/Ohm
- Charge Time: Less than 5 seconds to 200 Joules with a new/ fully charged lithium ion battery at 25o C
- Shock Delivery: Via multifunction defib electrode pads or paddles Shock-to-Shock Cycle Time: Typically less than 20 seconds seconds
- Minimum Patient Impedance Range: 15 Ohm (internal defibrillation); 25 Ohm (external defibrillation) Ohm
- Maximum Patient Impedance Range: 180 Ohm Ohm
- AED Mode: Shock advisory sensitivity and specificity meet AAMI DF-39 guide- lines
Environmental & physical requirements
- Solids/Water resistance: IP24
- Operating temperature: 0o – 45o C (32o – 113o F) ° C/° F
- Storage temperature: -20 – 70o C (-4o – 158o F) ° C/° F
- Operating humidity: 0% to 95% relative %
- Operating altitude: 0 to 15000 ft (0 to 4500 m) ft/m
- Storage altitude: 0 to 15000 ft (0 to 4500 m) ft/m
Battery
- Type: 6.3 Ah, 14.8 V, rechargeable lithium ion Ah/V
- Dimension: 6.5” (H) x 3.8” (W) x 1.6” (D) (165 mm x 95 mm x 42 mm) “/mm
- Weight: 1.6 lb. (0.73 kg) lb/kg
- Charge Time: Approximately 3 hours from fully depleted to 100%. 90 minutes from fully depleted to 80% hours/%
- Capacity(b): At least 3.5 hours of continuous 12-lead ECG/SpO2 and CO2 monitoring.
- With NBP every 15 minutes and pacing at 180 ppm at 160 mA on one new fully charged battery. hours