Mindray PM-9000 Express Features
- 12.1” Color TFT LCD Display
- Weight: ≤ 16.5 lbs.
- Batteries: Lithium or Sealed Lead-Acid
- Operating Time: 48 to 300 minutes
- Perfect fit for the ICU, CCU, Operating Room, Emergency Department, Observation Wards, Clinics, and hospital-type settings.
Mindray PM-9000 Express Specifications
Dimensions
- Height: 10.6” (27 cm)
- Width: 12.5” (31.8 cm)
- Depth: 5.4” (13.7 cm)
- Weight: 16.5 lbs (≤7.5 kg)
Display
- Type: Color TFT LCD
- Size: 12.1 inches (diagonal)
- Resolution: 800×600 pixels
Recorder
- Type: Thermal dot array
- Horizontal resolution: 160 dots/cm (at 25 mm/s recording rate)
- Vertical resolution: 80 dots/cm
- Width of the recorder paper: 50 mm
- Length of the recorder paper: 30 m
- Recording rate: 25 mm/s, 50 mm/s
- Recorded waveforms: 2
Data Storage
- Trend data: Long trend: 96 hours, resolution 1min, 5 min or 10 min.; Short trend: 1 hour, resolution 1 s or 5 s.
- Alarm events: 70 alarm events and associated waveforms (with user selectable waveform length 8s, 16 or 32).
- ARR events: 80 ARR events and associated waveforms with 8s wavelength.
- NIBP measurements: 800 NIBP groups, including systolic pressures, mean pressures, diastolic pressures and measurement time.
ECG
- Lead Type: 3, 5, or 12-Lead
- Measurement range: Neonate: 15 to 350 BPM; Pediatric: 15 to 350 BPM; Adult: 15 to 300 BPM
- Sensitivity selection: 1.25mm/mV (×0.125), 2.5mm/mV (×0.25), 5mm/mV (×0.5), 10mm/mV (×1), 20mm/mV (×2) and auto
- Sweep speed: 12.5mm/s, 25mm/s, 50mm/s
- Heart Rate Resolution: 1 BPM
- Heart Rate Accuracy: ±1BPM or ±1%, whichever is greater
- Sensitivity: 200μV (lead II)
- Arrhythmia Analysis: ASYSTOLE, VFIB/VTAC, PVC, COUPLET, VT>2, BIGEMINY, TRIGEMINY, R ON T, MISSED BEATS, TACHY, BRADY, PNC and PNP
RESP
- Measurement Technique: Thoracic impedance
- Lead: Optional: lead I and lead II; default lead II
- RR Measurement range: Adult: 0 to 120 BrPM; Pediatric/neonate: 0 to 150 BrPM
- RR Resolution: 1 BrPM
- RR Accuracy: 7 to 150 BrPM: ±2 BrPM or ±2%, whichever is greater
SpO2
- Compatible Modules: Mindray/Masimo/Nellcor
IBP
- Measurement Technique: Auto oscillation
- Displayed Parameters: Systolic pressure, diastolic pressure and mean pressure
- Mode of operation: Manual, auto and continuous
- Measurement interval in auto mode: 1/2/3/4/5/10/15/30/60/90/120/180/240/480 minutes
- Measurement time in continuous mode: 5 minutes
TEMP
- Number of channels: 2
- Displayed parameters: T1, T2 and TD
- Measurement range: 32 to 122°F (0 to 50°C)
- Resolution: 0.1°C
- Precision: 0.1°C (excluding the sensor): ±0.2°C (including the YSI 400 series sensor)
- Update period: 1s
CO
- Measurement Technique: Thermal dilution
- Calculated Parameter: CO, hemodynamics
- Measurement Range: CO: 0.1 to 20l/min; TB: 23 to 43°C; TI: 0 to 27°C
- Resolution: CO: 0.1 l /min; TB, TI: 0.1°C
CO2
- Measurement technique: Infrared absorption technique
- Measurement mode: Sidestream, microstream or mainstream (optional)
- Displayed parameter: EtCO2, FiCO2, Respiration Rate
- Compatible Modules: Mindary/Oridion/Welch Allyn
- CO2 measurement range: 0 to 99mmHg
AG
- Measurement technique: Infrared absorption
- Measurement mode: Side stream
- Gas Type: CO2, N2O, O2 (optional), Des, Iso, Enf, Sev, Hal