Midmark M11 UltraClave Features
- Intuitive display and simple prompts for easy operation
- Four pre-programmed sterilization cycles: Unwrapped, Pouched, Packs and Handpieces
- Dry Cycle Door Stop – Prevents door from opening full if there is a slight residual pressure in the chamber.
- Monitors chamber temperature during a cycle to prevent an overheat condition.
- Pressure relief valve opens to reduce chamber pressure if the chamber steam pressure exceeds the allowable limit.
Midmark M11 UltraClave Specifications
Dimensions
- Height (With Printer): 17.9 in. (45.2 cm)
- Width: 17.8 in. (45.2 cm)
- Depth (With Plug): 23.8 in. (60.5 cm)
- Counter Area: 17.8 x 21 in (45.2 x 53.3 cm)
- Chamber: 11.0 in diam. x 18 in deep (22.9 cm diam. x 46 cm deep)
- Large Tray: 9 in x 15 in x 1 1/8 in (22.9 cm x 38 cm x 2.9 cm)
- Small Tray: 6 5/8 in x 15 in x 1 1/8 in (14.3 cm x 38 cm x 2.9 cm)
- Weight (Empty): 99 lbs (44.9 kg)
- Weight (Full): 111 lbs (50.4 kg)
Water Reservoir Capacity
- Approximately 1.4 Gallons (5.3 liters) to full mark. Usable volume is 1 Gallon (3.8 liters)
Electrical Requirements
- 115 VAC, 50/60 Hz, 15 Amps Single Phase
- 115 VAC has 1425 watt tubular immersion heater
- 230 VAC, 50/60 Hz, 10 Amps Single Phase
- 230 VAC has 1500 watt tubular immersion heater
Chamber Pressure
- At 270 °F (132 °C) : 27.1 psi. (182 kPa)
- Safety Valve Setting: 40 psi. (275.8 kPa)
Heat Emission
- 5000 BTU/Hr during operation
Fuses
- 115 VAC models
- F1: 0.250 amp, 250 V, Slo-Blo, 1/4” x 1-1/4”
- F2: 15 amp, 250 V, Fast-Acting, 1/4” x 1-1/4”
- 230 VAC models:
- F1: 0.125 amp, 250 V, Slo-Blo, 5mm x 20mm
- F2: 8 amp, 250 V, Fast-Acting, 5mm x 20mm
Safety Standards
- UL Listed to U.S. & Canadian StandardsUL 61010-1, 2nd Edition
- IEC 61010-2-040, 1st Edition
- CAN/CSA C22.2 No. 61010-1, 2nd Edition
- CSA C22.2 No. 61010-2-040-07, Part 2-040, 1st Edition
- Meets the requirements of ASME Boiler and Pressure Vessel Code
- A separate (dedicated) circuit is recommended for this sterilizer