GE Datex Ohmeda E-CAiOv Module

Datex Ohmeda E-CaiOv Module tests CO2, N2O, O2, Anesthesia Agents, Agent Identification, and Spirometry. The E-CAioV Airway module is specifically designed for Anesthesia Monitor and Compact Anesthesia Monitor to measure anesthetic agent with identification, CO2, Patient Oxygen, Patient Spirometry™, and direct gas exchange.

Worldwide Sales & Service

  • Upto 50% Below OEM pricing
  • Reliable Equipment’s for your Facility

GE Datex Ohmeda E-CAiOv Module Features

  • All parameters measured at the patient’s airway
  • D-fend™ water separation system
  • Patient Oxygen – Datex-Ohmeda paramagnetic technology
  • Inspired and end-tidal values, inspired/expired difference, waveform
  • CO2 and N2O – Datex-Ohmeda infrared technology
  • Inspired and end-tidal values, waveform and respiration rate
  • Anesthetic agents – Datex-Ohmeda infrared technology
  • Measures and identifies all five agents: Halothane, Enflurane, Isoflurane, Sevoflurane and Desflurane

GE Datex Ohmeda E-CAiOv Module Specifications

Dimensions

  • Height: 4.4 in ( 112 mm)
  • Width:3 in (75mm)
  • Depth: 9 in (228 mm)
  • Weight: 3.5 lbs (1.6 kg)

Temperature

  • Operating temperature +10 to +40 °C
  • Storage temperature -25 to +70 °C
  • Atmospheric pressure
    • 666 to 1060 hPa /
    • (67 to 106 kPa)
    • (500 to 800 mmHg)
    • (666 to 1060 mbar)
  • Relative humidity 10 to 95% non-condensing (in airway 0 to 100%, condensing)

Power

  • Power consumption 12.6 W Prms, 14.6 W momentary
  • Protection against electrical shock Type BF

CO2 Performance

  • Measurement range 0 to 15 vol% (0 to 15 kPa, 0 to 113 mmHg)
  • Measurement rise time < 400 ms typical
  • Accuracy (0.2 vol% +2% of the reading)
    • e.g. Reading 5.0 % accuracy = ± (0.2 + 0.1); = ± 0.3 vol%
  • Gas cross effects < 0.2 vol% (O2, N2O, anesthetic agents)

O2 Performance

  • Measurement range 0 to 100 vol%
  • Measurement rise time < 400 ms typical
  • Accuracy ± (1 vol% + 2% of the reading)
  • Gas cross effects < 1 vol%; anesthetic agents
    • < 2 vol%; N2O
  • O2 Fi-Et difference resolution 0.1 vol%

N20 Performance

  • Measurement range 0 to 100%
  • Measurement rise time < 400 ms typical
  • Accuracy ± (2 vol% + 2% of the reading) (0%
    • ± (2 vol% + 8% of the reading) (85%
  • Gas cross effects < 2 vol%; anesthetic agents

Respiration Rate

  • Measurement range 4 to 60 breaths/min
  • Detection criteria 1% variation in CO2

Additional information

Manufacturer

GE