Cardiocap 5 Features
The Cardiocap/5 is a compact, all-in-one monitor that offers comprehensive vital signs monitoring capabilities for all anesthesia monitoring locations throughout the perioperative environment. Its compact size fits in places where space is at a premium. The frame integrated mounting piece allows quick and easy mounting on wall rail, roll stand, and anesthesia machine mounting arms.
The Cardiocap/5 is designed for use in the operating room, ambulatory surgery unit, induction room, and PACU. The Cardiocap/5 offers full vital signs monitoring capabilities from the monitoring of oxygenation and circulation to advanced airway gas analysis and ventilation measurement with patient spirometry. Configuration options allow you to choose the set of parameters that best suit your needs.
- 10.4” color LCD display.
- Recorder that prints trend data and record parameter waveforms.
- Hemodynamic and airway gas monitor specifically designed for operating and recovery rooms.
- Dedicated anesthesia menu logic.
- Small and Compact.
- Built-in back-up battery.
- Convenient mounting system.
Cardiocap 5 Specifications
Dimensions
- Height: 11.8” (30 cm)
- Width: 13” (33 cm)
- Depth: 8.7” (22 cm)
- Weight: <24.8 lbs (<11.2 kg)
LCD Display
- Type: Active matrix color LCD
- Size (Diagonal): 10.4”
- Resolution: 640 x 480
Battery
- Type: 12V 2.6AH, lead acid
- Back-up battery time: at least 15 minutes when fully charged
- Charging time: 5 hours (typical)
- The green battery charge status LED is On when the battery is fully charged, on the holding voltage. The LED flashes when the battery is being charged.
NIBP
- Measurement range:
- Adult 25 to 260 mmHg (3.3 to 34.7 kPa)
- Child 25 to 195 mmHg (3.3 to 26.0 kPa)
- Infant 15 to 145 mmHg (2.0 to 19.3 kPa)
- Pulse Rate Range Accepted: 30 to 250 bpm
- Typical Measuring Time: Adults 23 seconds, infants 20 seconds
Temperature
- Measurement Range: 50 to 113°F (10 to 45°C)
- Measurement Accuracy: 77 to 113 °F ± 0.2 °F (25 to 45.0 °C ± 0.1 °C ); 50 to 76.8 °F ± 0.4 °F (10 to 24.9 °C ± 0.2 °C)
- Probe Type: Compatible with Datex-Ohmeda temperature probes only
ECG
- Waveform display (with 50 Hz power supply frequency):
- Monitoring filter: 0.5 to 30 Hz
- ST filter: 0.05 to 30Hz
- Diagnostic filter: 0.05 to 100 Hz
- Waveform display (with 60 Hz power supply frequency):
- Monitoring filter: 0.5 to 40 H
- ST filter: 0.05 to 40 Hz
- Diagnostic filter: 0.05 to 100 Hz
- Heart rate
- Measurement range: 30 to 250 bpm
- Measurement accuracy: ± 5% or ± 5
- Pacemaker pulse detection level: 2 to 500 mV
- Pacemaker pulse duration: 0.5 to 2 ms
Impedance Respiration
- Respiration Range: 4 to 120 respirations/minute
- Accuracy: ± 5% or ± 5 bpm
Pulse Oximetry (Standard)
- Display Update Time: 5 seconds
- Averaging Time: Adjustable
- Plethysmographic Waveform Scaling: Adjustable
- SpO2
- Calibration range: 50 to 100%
- Calibrated against functional saturation
- Measurement range: 40 to 100%
- Measurement accuracy (% SpO2 ±1 SD):
- 80 to 100% ± 2 digits;
- 50 to 80% ± 3 digits;
- Below 50% unspecified
- NOTE: SpO2 measurement accuracyis based on deep hypoxia studies using Datex-Ohmeda FingerSat sensors on volunteered subjects. Arterial blood samples were analyzed by a Radiometer OSM CO-oximeter. Refer to the sensor instructions for specific SpO2 accuracy data.
- Pulse rate
- Measurement range: 30 to 250 bpm
- Measurement accuracy: ± 5% or ± 5 bpm
- Default alarm limits
- SpO2: high Off, low 90%
- Pulse rate: high 160, low 40
- NOTE: Limits are adjustable.
- Sensor emitter wavelength ranges
- Red LED: 660 nm
- Infrared LED: 900 nm
Options
- Pulse Oximetry, Datex-Ohmeda Enhanced (N-XOSAT)
- Pulse Oximetry, Nellcor Compatible (N-XNSAT)
- Invasive Blood Pressure ( N-XP)
- Airway Gasses (N-XC, N-XCO, and N-XCAiO): CO2, O2, N2O, Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane
- Patient Spirometry (N-XV)
- NeuroMuscular Transmission (N-XNMT)
- Recorder (N-XREC)