Drager Vamos Features
- Inferred light measurement from sidestream samples.
- Measured Parameters: CO2, N2O, SpO2, and 5 anesthetic agents.
- Gas measurement: breath to breath resolution.
- High-contrast amber electroluminescent display.
- 5” diagonal display.
Drager Vamos Specifications
Dimensions
- Height: 6.5” (16.6 cm)
- Width: 9.4” (24 cm)
- Depth: 6.5” (16.6 cm)
- Weight: 5.3 lbs. (2.4 kg)
Display
- Type: High-contrast amber electroluminescent display with excellent vertical and horizontal viewing angles
- Display Size: 5” diagonal 4.25” x 2.28” (127 mm diagonal 108 x 58 mm)
- Graphic Display of CO2: waveform
- Numeric Display: Fractional inspiratory concentration (Fi) of CO2, N2O and of selected agent (Vamos), End-tidal concentration (Fet) of CO2, N2O and selected agent (Vamos)
CO2
- Sidestream Sample Flow Rate: 150 mL/min ±20 mL/min
- Measuring Range: 0 to 10.0 vol.% / kPa, 0 to 75 mmHg
- Prevision: ±0.5% Vol.% / kPa or ± 12 % rel. which ever is greater; ±4 mmHg or ±12% rel., whichever is greater
- Resolution: 1 mmHg / 0.1 Vol.% / 0.1 kPa
Anesthetic Gas Measurement
- Sidestream Samle Flow Rate: 150 mL/min ±20 mL/min
- Response Time: <500 ms
- Display Range for N2O: 0 to 99% vol.%
- Display Range for Halothane: 0 to 8.5 vol.%
- Display Range for Isoflurane: 0 to 8.5 vol.%
- Display Range for Enflurane: 0 to 10 vol.%
- Display range for Sevoflurane: 0 to 10 vol.%
- Display Range for Desflurane: 0 to 22 vol.%
- Warm-up Phase: 4 minutes
SpO2
- Oxygen Saturation SpO2: 1 to 100%
- Resolution: 1%
- Pulse Rate: 20 to 250/min
- Resolution: 1/min
Interfaces
- RS 232 C
- MEDIBUS RS 232 C