Drager Fabius MRI

The Drager Fabius MRI is an anesthesia machine that is compatible with an MRi Machines. I can be used with MRI Systems of 1.5 and 3 Tesla.The Fabius MRI offers a variety of ventilation modes including volume and pressure. The Drager MRI Anesthesia Machine comes with an E-vent piston controlled ventilator – this can deliver a maximum peak flow of ventilation and does not require a drive gas. The anesthesia Machine has a large, high-visibility color screen along with an integrated warning and alarm LED system.

Worldwide Sales & Service

  • Upto 50% Below OEM pricing
  • Reliable Equipment’s for your Facility

Drager Fabius MRI Features

  • High visibilty color screen.
  • Integrated Warning and alarm LEDs.
  • 2 vaporizer mount.
  • Pressure guages for pin index gas cylinders.
  • Dräger Sob CLIC absorber.
  • Anesthetic Gas Scavenging System.
  • 3 Large drawers, 1 with lock.
  • Central brake.
  • Compact Breathing System.
  • E-vent piston Ventilator.

Dräger Fabius MRI Specifications

Base Unit

  • Trolley Version (Cart) with COSY: Dimensions (W × H × D): approximately 39 × 55 × 35.5 in. (99 × 140 × 90 cm)
  • Fabius MRI Trolley (with COSY) Weight and load without supplementary cylinders and vaporizers: 365 lbs. (165.8 kg)

Power and Battery Backup

  • Power Input: 100 to 240 VAC, 50 / 60 Hz, 70 VA, including additional power outlets.
  • Operation time with fully charged batteries: > 45 min.

Anesthesia Gas Supply Module

  • Range of fresh gas flow indicators: 0.0 to 12.0 L/min Total fresh gas flowmeter 0 to 10 L/min.
  • O2 flush: at 87 psi (6 bar): max 75 L/min; at 41 psi (2.8 bar): min. 25 L/min.
  • Vaporizer: 2 position Dräger mount (Interlock 2 – System): Dräger Isoflurane Vapor 2000, Dräger Sevoflurane Vapor 2000, Dräger Halothane Vapor 2000, Dräger Enflurane Vapor 2000.

Ventilator Operating Specifications

  • Ventilator E-vent: Electronically controlled, electrically driven.
  • Operating modes: Volume Controlled Ventilation, Pressure Controlled Ventilation, Pressure Support, SIMV/PS, Manual Ventilation, Spontaneous Breathing.

Control Input Ranges

  • Breathing Frequency (rate): 4 to 60 bpm.
  • Positive End Expiratory Pressure (PEEP): 0 to 20 cmH2O (hPa).
  • Inspiration/expiration ratio (Ti:Te): 4:1 to 1:4.
  • Pressure limiting (Pmax): 15 to 70 cmH2O (hPa).
  • Tidal Volume (Vt): 20 to 1400 mL in Volume Control, 20 to 1100 mL in SIMV/PS.
  • Inspiration pause (Tip:Ti): 0 to 50 %.
  • SIMV Inspiratory time: 0.3 – 4.0 sec.
  • Inspiratory pressure (Pinsp): PEEP + 5 to 65 cmH2O (hPa).
  • Inspiratory Flow (InspFlow): 10 to 75 L/min in Volume and Pressure Control modes, 10 to 85 L/min in Pressure Support and SIMV/PS modes.
  • Pressure Support Level (PPS): PEEP + 3 to 20 cmH2O (hPa) Min. frequency for apnoe-ventilation (Freq. Min.) 3 to 20 bpm and “OFF”.

Additional information

Manufacturer

Drager