Covidien Capnostream 20p

The Capnostream 20p patient monitor is built on a legacy of proven performance. For nearly two decades, clinicians have relied on Microstream technology for an accurate, continuous view of ventilation adequacy on intubated and non-intubated patients, from neonate to adult. The Capnostream 20p monitor, with the Apnea-Sat Alert feature, alerts caregivers to repetitive patterns of apneas and oxygen desaturation, revealing a more complete picture of a patient’s respiratory status.

Worldwide Sales & Service

  • Upto 50% Below OEM pricing
  • Reliable Equipment’s for your Facility

Covidien Capnostream 20p Features

  • Supports current standards of care by providing etCO2 and SpO2 measurements.
  • The Capnostream 20p patient monitor with Microstream includes Smart Alarm Management Integrated Algorithms, a suite of algorithms proven to reduce alarms and simplify the use of patient monitoring to enhance patient safety and improve clinical efficiency.
  • Screens out low amplitude ‘non-breath’ etCO2 excursions like, talking, crying or snoring.
  • Reduces clinically insignificant alarms.
  • Alerts clinicians to clinically significant changes in pulse oximetry.
  • IPI algorithm presents one value that demonstrates real-time respiratory status based on etCO2, RR, PR and SpO2.
  • Tracks and reports apneas per hour and oxygen desaturation index, and provides the data in trend reports.
  • Compatible with central monitoring systems and interfaces with nurse call systems and electronic medical records.
  • Provides convenient bedside documentation.

Covidien Capnostream 20p Specifications

Dimensions

  • Height: 6.6″ (16.7 cm)
  • Width: 8.7″ (22 cm)
  • Depth: 7.6″ (19.2 cm)
  • Weight: 7.94 lbs (3.6 kg)

Display

  • Type: LCD with chemically etched glass layer
  • Diagonal: 12.1″ (30.7 cm)

Power

  • Power Supply Input Voltage 100-240 V, 50/60 Hz, ± 10%.

Microstream Capnography

  • Units: mmHg, kPa or Vol%.
  • Measurement range: etCO2, FiCO2 0-150 mmHg.
  • Respiratory Rate: 0 bpm-150 bpm.
  • Resolution: 1 mmHg.
  • Accuracy (±SD): etCO2, FiCO2 0-38 mmHg ± 2 mmHg. 39-150 mmHg ± (5% of reading + 0.08% for every 1 mmHg above 38 mmHg).
  • Respiratory rate: 0-70 bpm ±1 bpm, 71-120 bpm ±2 bpm, 121-150 bpm ±3 bpm.
  • Flow rate: 50 ml/minute, flow measured by volume.
  • Waveform sampling: 50 millisecond intervals.
  • Response time: 2.95 second (typical).
  • Initialization time: 40 second (typical).
  • Calibration interval: Initial calibration after 1,200 operating hours, then annually or after 4,000 operating hours, whichever comes first.

Nellcor Pulse Oximetry

  • Measurement range 0-100%.
  • Accuracy
    • Adult mode: SpO2 range 70%-100% ± 2 digits; SpO2 range 0-69% unspecified.
    • Neonate mode: SpO2 range 70%-100% ± 3 digits; SpO2 range 0-69% unspecified.
    • Pulse rate: Range: 20-250 BPM; Accuracy: ± 3 BPM; Sat Sec Range: 10, 25, 50, 100.

Additional information

Manufacturer

Covidien